FDA Device recall Z-0318-2026
Brand Name: Dexcom G6 Continuous Glucose Monitoring System Product Name: Dexcom G6 and G6 Pro Android US CGM App Model/Catalog Number: SW11…
- FDA Device
- Class I
- ongoing
- Published 2025-11-05
On 2025-11-05 the FDA classified a Class I recall of Brand Name: Dexcom G6 Continuous Glucose Monitoring System Product Name: Dexcom G6 and G6 Pro Android US CGM App Model/Catalog Number: SW11… by Dexcom, citing A software defect in version v1.15.0 of the G6 Android app can cause the app to terminate unexpectedly, which may result in the user not receiving estimated glucose values, alarms….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-0318-2026 |
| Classification | Class I |
| Status | ongoing |
| Initiated | 2025-08-28 |
| Published | 2025-11-05 |
| Last updated | 2026-09-13 11:57 UTC |
| Company | Dexcom |
| Distribution | US |
Product
Brand Name: Dexcom G6 Continuous Glucose Monitoring System Product Name: Dexcom G6 and G6 Pro Android US CGM App Model/Catalog Number: SW11678 Software Version: version 1.15.0 Product Description: Dexcom G6 and G6 Pro Android US CGM App SW11678 version 1.15.0 Component: Dexcom G6 Continuous Glucose Monitoring System
Reason
A software defect in version v1.15.0 of the G6 Android app can cause the app to terminate unexpectedly, which may result in the user not receiving estimated glucose values, alarms, alerts or notifications. This could result in the missed detection of a hyperglycemic or hypoglycemic event, protentional resulting in severe hyperglycemia, diabetic ketoacidosis (DKA), or hyperosmolar hyperglycemic state (HHS).
Identifiers
| code_info | Lot Code: UDI: SW11678 - 00386270000811 |
|---|
Status history
No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:57 UTC. Current status: ongoing.
Source
Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-0318-2026%22