FDA Device recall Z-0317-2025
The FLUOROspeed is intended to be used as a universal diagnostic imaging system for radiographic and fluoroscopic studies
- FDA Device
- Class II
- ongoing
- Published 2024-11-13
On 2024-11-13 the FDA classified a Class II recall of The FLUOROspeed is intended to be used as a universal diagnostic imaging system for radiographic and fluoroscopic studies by Shimadzu Medical Systems Division, citing X-ray R/F system software with certain versions of the control board may occasionally fail to display images when switching from Fluoroscopy mode to Radiography mode, which may le….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-0317-2025 |
| Classification | Class II |
| Status | ongoing |
| Initiated | 2024-09-06 |
| Published | 2024-11-13 |
| Last updated | 2026-09-13 11:57 UTC |
| Company | Shimadzu Medical Systems Division |
| Distribution | US |
Product
The FLUOROspeed is intended to be used as a universal diagnostic imaging system for radiographic and fluoroscopic studies. Using a flat detector, it can perform a range of applications including general R/F, angiography and pediatric examinations. FLUOROspeed may be used for emergency treatment on an outpatient basis, as well as for bedside examinations.
Reason
X-ray R/F system software with certain versions of the control board may occasionally fail to display images when switching from Fluoroscopy mode to Radiography mode, which may lead to the possibility of delay or cancellation of the examination.
Identifiers
| code_info | UDI-DI: 04540217065271. Serial Numbers: MQ493E3E3001, MQ493E3E3002, MQ493E3D6002, MQ493E3D5001, MQ493E3D8002, MQ493E3E…UDI-DI: 04540217065271. Serial Numbers: MQ493E3E3001, MQ493E3E3002, MQ493E3D6002, MQ493E3D5001, MQ493E3D8002, MQ493E3E1002, MQ493E3D4002, MQ493E3D6003, MQ493E3DA002, MQ493E3E2001, MQ493E3E2002, MQ493E3E1001, MQ493E3DC001, MQ493E3D9001, MQ493E3D4005, MQ493E3D7001, MQ493E3DB002, MQ493E3D2002, MQ493E3D4001, MQ493E3D7002, MQ493E3D4003, MQ493E3D4004, MQ493E3D2001, MQ493E3D8001, MQ493E3D5002, MQ493E3DB001, MQ493E3D2003, MQ493E3DA001, MQ493E3D6001, MQ493E3E1003 |
|---|
Status history