FDA Device recall Z-0314-2017
SimpliRED D-Dimer, Product Code DSRK4 The SimpliRED D-dimer assay is a rapid, qualitative test for the detection of cross-linked fibrin deg…
- FDA Device
- Class II
- terminated
- Published 2016-11-02
On 2016-11-02 the FDA classified a Class II recall of SimpliRED D-Dimer, Product Code DSRK4 The SimpliRED D-dimer assay is a rapid, qualitative test for the detection of cross-linked fibrin deg… by Bbi Solutions Oem, citing A component of the kit is showing a sensitivity performance detected in stability testing at 15 months of an 18-month shelf life. Qualitative results may be affected. False negati….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-0314-2017 |
| Classification | Class II |
| Status | terminated |
| Initiated | 2016-06-01 |
| Published | 2016-11-02 |
| Last updated | 2026-09-13 11:54 UTC |
| Company | Bbi Solutions Oem |
| Distribution | CT, IN |
Product
SimpliRED D-Dimer, Product Code DSRK4 The SimpliRED D-dimer assay is a rapid, qualitative test for the detection of cross-linked fibrin degradation products containing the cross-linked D-diner site in human whole blood
Reason
A component of the kit is showing a sensitivity performance detected in stability testing at 15 months of an 18-month shelf life. Qualitative results may be affected. False negative results are possible which may incorrectly influence the clinical diagnostic decision.
Identifiers
| code_info | Lot Numbers containing test reagent lot # 1012-34437): BU183A, BU183B, BU184A, BU185A, BU185B Expiration Date: 31 JUL 2016 |
|---|