RecallRadar

FDA Device recall Z-0313-2017

VITROS Immunodiagnostic Products HBsAg Controls, Product Code 6800598 For in vitro diagnostic use only

On 2016-11-02 the FDA classified a Class III recall of VITROS Immunodiagnostic Products HBsAg Controls, Product Code 6800598 For in vitro diagnostic use only by Ortho Clinical Diagnostics, citing the signal/cutoff (s/c) results may increase throughout the open-reconstituted timeframe of the allowed 5 days. Increased results may occur on both the negative and positive Contr….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-0313-2017
ClassificationClass III
Statusterminated
Initiated2016-06-22
Published2016-11-02
Last updated2026-09-13 11:54 UTC
CompanyOrtho Clinical Diagnostics
DistributionUS

Product

VITROS Immunodiagnostic Products HBsAg Controls, Product Code 6800598 For in vitro diagnostic use only. For use in monitoring the performance of the VITROS ECi/ECiQ Immunodiagnostic Systems, the VITROS 3600 Immunodiagnostic System and the VITROS 5600 Integrated System when used for the in vitro qualitative detection of Hepatitis B Surface Antigen (HBsAg) in human serum and plasma (heparin, EDTA or citrate). The performance of the VITROS Immunodiagnostic Products HBsAg Controls has not been established with any other HBsAg assay.

Reason

The signal/cutoff (s/c) results may increase throughout the open-reconstituted timeframe of the allowed 5 days. Increased results may occur on both the negative and positive Controls. The positive Control, although showing elevated results, was still within acceptable performance.

Identifiers

code_infoProduct Code 6800598-Lot #'s: 1) 0860, 2) 0870, 3) 0880, 4) 0890, 5) 0908 Expiration: 1) 29-MAR-17, 2) 12-JUN-17, 3) 02-JUL-17, 4) 23-NOV-17, 5) 29-JAN-18

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:54 UTC. Current status: terminated.

Source

Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-0313-2017%22

RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.