FDA Device recall Z-0308-2022
Abbott HeartMate Touch Communication System, REF: HMT1100 (US), HMT1100-R (Rental US), HMT1150 (EU)
- FDA Device
- Class II
- ongoing
- Published 2021-12-08
On 2021-12-08 the FDA classified a Class II recall of Abbott HeartMate Touch Communication System, REF: HMT1100 (US), HMT1100-R (Rental US), HMT1150 (EU) by Thoratec, citing if LVAS communication system is trying to establish Bluetooth connection and another Bluetooth-enabled device is nearby and advertising for Bluetooth connection, the Bluetooth con….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-0308-2022 |
| Classification | Class II |
| Status | ongoing |
| Initiated | 2021-10-22 |
| Published | 2021-12-08 |
| Last updated | 2026-09-13 11:56 UTC |
| Company | Thoratec |
| Distribution | US |
Product
Abbott HeartMate Touch Communication System, REF: HMT1100 (US), HMT1100-R (Rental US), HMT1150 (EU)
Reason
If LVAS communication system is trying to establish Bluetooth connection and another Bluetooth-enabled device is nearby and advertising for Bluetooth connection, the Bluetooth connectivity interference may cause the system App to close or fail to open. Once Bluetooth adaptor is connected, barring disconnection, all functionality is unaffected; the device will continue to operate as expected.
Identifiers
| code_info | All serial numbers. |
|---|
Status history
No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:56 UTC. Current status: ongoing.
Other recalls by Thoratec
- Brand Name: HeartMate Product Name: HeartMate 3" System Controller Model/Catalog Number: 106531US Software Version: N/A Product Description…
FDA Device · Class II · 2026-06-24 · Thoratec - Brand Name: HeartMate Product Name: HeartMate 3 Left Ventricular System (LVAS) Kits Model/Catalog Number: 106524US Software Version: N/A Pr…
FDA Device · Class II · 2026-06-24 · Thoratec - Brand Name: HeartMate Product Name: HM 11V Li-ion Backup Battery, ABT Model/Catalog Number: 106128 Software Version: N/A Product Descriptio…
FDA Device · Class II · 2026-06-24 · Thoratec - CentriMag Blood Pump, REF: 102953, 201-20003, 201-90010, 201-90016
FDA Device · Class II · 2026-01-21 · Thoratec - Brand Name: HeartMate Product Name: HeartMate 3 Controller (Standalone) Model/Catalog Number: 106531US, HeartMate 3 Controller, US
FDA Device · Class II · 2025-11-26 · Thoratec
Source
Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-0308-2022%22