RecallRadar

FDA Device recall Z-0307-2022

The Ocean Water Seal Chest Drain is a plastic, three bottle vacuum collection device

On 2021-12-01 the FDA classified a Class II recall of The Ocean Water Seal Chest Drain is a plastic, three bottle vacuum collection device by Atrium Medical, citing inadequate/inappropriate Set-Up Instructions, which may result in delay in treatment and an unreasonable risk of harm to the patient.

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-0307-2022
ClassificationClass II
Statusongoing
Initiated2021-10-22
Published2021-12-01
Last updated2026-09-13 11:56 UTC
CompanyAtrium Medical
DistributionUS

Product

The Ocean Water Seal Chest Drain is a plastic, three bottle vacuum collection device. It consists of a collection chamber that holds shed mediastinal blood, a water seal chamber using a water seal that prevents air from entering into the pleural space and a suction control chamber, which incorporates a mechanical regulator to control suction pressures imposed upon the patient's pleural cavity.

Reason

Inadequate/inappropriate Set-Up Instructions, which may result in delay in treatment and an unreasonable risk of harm to the patient

Identifiers

code_info
Product Code/REF Number 2002-400; Lot codes: 441533, 443300, 443608, 443629, 444473, 445527, 448966, 449683, 450899, 45…Product Code/REF Number 2002-400; Lot codes: 441533, 443300, 443608, 443629, 444473, 445527, 448966, 449683, 450899, 450900, 451270, 451988, 452177, 452814, 454577, 455871, 458264, 458604, 459198, 459642, 462192, 462194, 464075, 464076, 465037, 466247, 469921, 470627, 471651,471703, 472562, 472785 and 473725. UDI Code: 00650862100211

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:56 UTC. Current status: ongoing.

Source

Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-0307-2022%22

RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.