Artis Q ceiling with software VD12-angiography systems developed for single and biplane diagnostic imaging and interventional procedure Mod…

On 2021-12-01 the FDA classified a Class II recall of Artis Q ceiling with software VD12-angiography systems developed for single and biplane diagnostic imaging and interventional procedure Mod… by Siemens Medical Solutions Usa, citing if SID (source-to-image distance) lift movement is activated and x-ray shall be released simultaneously; x-ray is not possible and the message No x-ray, try again is displayed, ma….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-0301-2022
ClassificationClass II
Statusterminated
Initiated2021-10-20
Published2021-12-01
Last updated2026-09-13 11:56 UTC
CompanySiemens Medical Solutions Usa
DistributionUS

Product

Artis Q ceiling with software VD12-angiography systems developed for single and biplane diagnostic imaging and interventional procedure Model Number: 10848281

Reason

If SID (source-to-image distance) lift movement is activated and x-ray shall be released simultaneously; x-ray is not possible and the message No x-ray, try again is displayed, may result in a short delay in procedure

Identifiers

code_info
Serial Numbers: 110088 110059 110103 110035 110046 110005 110058 110056 110062 109523 109645 110067 110095…Serial Numbers: 110088 110059 110103 110035 110046 110005 110058 110056 110062 109523 109645 110067 110095 110051 110001 110020 110007 110093 110069 110075 109627 110042 110043 110048 110061 110063 110064 110032 110047 110102 110039 110040 110023 110065 110052 110018 110072 110081 110036 110054 110053 110044 110031 110016 110037 110077 109497 110079 110099 110027 110049 110066 110024 110017 110025 110100 110083 110091 110006 110055 110026 110022 110000 133501 110057 110104 110041 Expanded Recall 11/15/21: 110096 110014