FDA Device recall Z-0298-2019
AXIOM Multix MT (model number 8395415 or 8395399 or 8395381), a patient table used in conjunction with the Siemens Multix PRO- and TOP Syst…
- FDA Device
- Class II
- terminated
- Published 2018-11-07
On 2018-11-07 the FDA classified a Class II recall of AXIOM Multix MT (model number 8395415 or 8395399 or 8395381), a patient table used in conjunction with the Siemens Multix PRO- and TOP Syst… by Siemens Medical Solutions Usa, citing in rare cases of high clinical workload, the steel cables inside the lifting column of the overhead tube, which are designed to take the load, may rapture without triggering the s….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-0298-2019 |
| Classification | Class II |
| Status | terminated |
| Initiated | 2018-08-06 |
| Published | 2018-11-07 |
| Last updated | 2026-09-13 11:54 UTC |
| Company | Siemens Medical Solutions Usa |
| Distribution | AL, AR, AZ, CA, CO, DC, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MI, MN, MO, MS, NC, ND, NE, NM, NV, NY, OH, OK, PA, PR, SC, SD, TX, UT, VA, WI, WY |
Product
AXIOM Multix MT (model number 8395415 or 8395399 or 8395381), a patient table used in conjunction with the Siemens Multix PRO- and TOP Systems radiographic exposures of the entire body.
Reason
In rare cases of high clinical workload, the steel cables inside the lifting column of the overhead tube, which are designed to take the load, may rapture without triggering the safety lock. This could lead to overloading the safety rope, which is not designed for continuous load and ongoing movement under load. This may lead to mechanical fatigue and cause the arm to drop down unexpectedly during movement in vertical direction and during patient positioning.
Identifiers
| code_info | Serial Numbers 1137 1138 1525 1121 1132 1529 1074 1531 1472 1136 1119 |
|---|
Status history
No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:54 UTC. Current status: terminated.
Source
Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-0298-2019%22