FDA Device recall Z-0290-2024
Cardiohelp-i, blood oxygenation and carbon dioxide removal system, Part Numbers 701072780 (Cardiohelp-i), 701068550 (Slide Rail Holder)
- FDA Device
- Class II
- ongoing
- Published 2023-11-15
On 2023-11-15 the FDA classified a Class II recall of Cardiohelp-i, blood oxygenation and carbon dioxide removal system, Part Numbers 701072780 (Cardiohelp-i), 701068550 (Slide Rail Holder) by Maquet Medical Systems Usa, citing it was found that the fixation (weld) employed by the supplier of the nut bolt to the slide rail plate was insufficient. The deficient weld allows the nut bolt to break off from t….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-0290-2024 |
| Classification | Class II |
| Status | ongoing |
| Initiated | 2023-10-12 |
| Published | 2023-11-15 |
| Last updated | 2026-09-13 11:56 UTC |
| Company | Maquet Medical Systems Usa |
| Distribution | US |
Product
Cardiohelp-i, blood oxygenation and carbon dioxide removal system, Part Numbers 701072780 (Cardiohelp-i), 701068550 (Slide Rail Holder)
Reason
It was found that the fixation (weld) employed by the supplier of the nut bolt to the slide rail plate was insufficient. The deficient weld allows the nut bolt to break off from the slide rail plate with slight manual force.
Identifiers
| code_info | UDI-DI: 04058863074863; Serial Numbers: 90414355 through 90414868 |
|---|
Status history
No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:56 UTC. Current status: ongoing.
Source
Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-0290-2024%22
RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.