RecallRadar

FDA Device recall Z-0289-2017

Vari-Safe Injection Needle US endoscopy 5976 Heisley Road Mentor, Ohio 44060 A sterile, single use, flexible sheath device intended to be u…

On 2016-10-26 the FDA classified a Class II recall of Vari-Safe Injection Needle US endoscopy 5976 Heisley Road Mentor, Ohio 44060 A sterile, single use, flexible sheath device intended to be u… by Us Endoscopy Group, citing US Endoscopy is conducting a voluntary product recall of one lot of the Vari-Safe Injection Needle as it was identified that the lot was incorrectly distributed. The lot was assem….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-0289-2017
ClassificationClass II
Statusterminated
Initiated2016-05-18
Published2016-10-26
Last updated2026-09-13 11:54 UTC
CompanyUs Endoscopy Group
DistributionOH, OR, WA

Product

Vari-Safe Injection Needle US endoscopy 5976 Heisley Road Mentor, Ohio 44060 A sterile, single use, flexible sheath device intended to be used for the injection of various types of media through flexible endoscopes

Reason

US Endoscopy is conducting a voluntary product recall of one lot of the Vari-Safe Injection Needle as it was identified that the lot was incorrectly distributed. The lot was assembled with internal catheter components which differ from the original Vari-Safe needle internal catheter components. All components are dimensionally similar to the original Vari-Safe needle internal catheter components and the components have history of use on other US Endoscopy needle products.

Identifiers

code_infoModel #: 00711819; Lot 1520268

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:54 UTC. Current status: terminated.

Source

Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-0289-2017%22

RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.