FDA Device recall Z-0286-2014
The Plum A+3 Infusion System is a cassette based multifunction infusion system that allows two lines in and one line out and can be used fo…
- FDA Device
- Class II
- terminated
- Published 2013-11-20
On 2013-11-20 the FDA classified a Class II recall of The Plum A+3 Infusion System is a cassette based multifunction infusion system that allows two lines in and one line out and can be used fo… by Hospira, citing plum A+ infusers have the potential for the distal (occlusion) press sensor pin to break. The distal pressure sensor pin is a part of the overall subsystem that measures the press….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-0286-2014 |
| Classification | Class II |
| Status | terminated |
| Initiated | 2013-02-01 |
| Published | 2013-11-20 |
| Last updated | 2026-09-13 11:53 UTC |
| Company | Hospira |
| Distribution | US |
Product
The Plum A+3 Infusion System is a cassette based multifunction infusion system that allows two lines in and one line out and can be used for standard, piggyback, or concurrent delivery. The Plum A+3 Infusion System is designed to deliver parenteral, enteral, or epidural infusions over a broad range of infusion rates from multiple fluid container types. The Plum A+3 Infusion System includes an infuser, disposable IV sets (hereafter called a set), optional accessories, and this operator's manual. The Plum A+ Infusion System with Hospira MedNet Software is indicated for use in parenteral, enteral and epidural therapies and the administration of whole blood and blood products.
Reason
Plum A+ infusers have the potential for the distal (occlusion) press sensor pin to break. The distal pressure sensor pin is a part of the overall subsystem that measures the pressure within the distal line of the administration set and indicates the presence of a full or partial distal occlusion. A broken distal pin can only be detected via a visual inspection of the distal pressure pin. A broken
Identifiers
| code_info | 1) List Number 12348; Serial Numbers: 13740156 , 13740256 , 13740266 , 13740351 , 13740380 , 13740451 , 13740508…1) List Number 12348; Serial Numbers: 13740156 , 13740256 , 13740266 , 13740351 , 13740380 , 13740451 , 13740508 , 13740612 , 13740632 , 13740731 , 13740732 , 13740733 , 13740734 , 13740735 , 13740736 , 13740737 , 13740738 , 13740739 , 13740740 , 13740741 , 13740742 , 13740743 , 13740744 , 13740745 , 13740746 , 13740747 , 13740748 , 13740749 , 13740750 , 13740751 , 13740752 , 13740753 , 13740754 , 13740755 , 13740756 , 13740757 , 13740758 , 13740759 , 13740760 , 13740761 , 13740762 , 13740763 , 13740764 , 13740765 , 13740766 , 13740767 , 13740768 , 13740769 , 13740770 , 13740771 , 13740772 , 13740773 , 13740774 , 13740775 , 13740776 , 13740777 , 13740778 , 13740779 , 13740780 , 13740781 , 13740782 , 13740783 , 13740784 , 13740785 , 13740786 , 13740787 , 13740788 , 13740789 , 13740790 , 13740791 , 13740792 , 13740793 , 13740794 , 13740795 , 13740796 , 13740797 , 13740798 , 13740799 , 13740800 , 13740801 , 1 |
|---|