Continuous CT Fluoroscopy DVI Ceiling Mount - CCT Arm DVI Option - Gemini TF 16 System Number: 882470 Continuous CT (CCT) is a CT or PET/CT…

On 2021-12-01 the FDA classified a Class II recall of Continuous CT Fluoroscopy DVI Ceiling Mount - CCT Arm DVI Option - Gemini TF 16 System Number: 882470 Continuous CT (CCT) is a CT or PET/CT… by Philips North America, citing significant wear of the Monitor Bracket Assembly shaft may go unrecognized could lead to the Monitor Bracket Assembly detaching from the Suspension Arm, and resulting in injury to….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-0280-2022
ClassificationClass II
Statusongoing
Initiated2021-05-01
Published2021-12-01
Last updated2026-09-13 11:56 UTC
CompanyPhilips North America
DistributionUS

Product

Continuous CT Fluoroscopy DVI Ceiling Mount - CCT Arm DVI Option - Gemini TF 16 System Number: 882470 Continuous CT (CCT) is a CT or PET/CT product option that enables physicians and radiologists to produce real-time CT images to facilitate image-guided surgery.

Reason

Significant wear of the Monitor Bracket Assembly shaft may go unrecognized could lead to the Monitor Bracket Assembly detaching from the Suspension Arm, and resulting in injury to patient, operator, bystander, or service person located in the path of a falling part

Identifiers

code_infoSerial Number: 7084 7145 7584