RecallRadar

FDA Device recall Z-0278-2014

Medtronic RestoreUltra, model 37712, and RestoreSensor, model 37714, Multi-program Rechargeable Neurostimulator for Spinal Cord Stimulation

On 2013-11-20 the FDA classified a Class II recall of Medtronic RestoreUltra, model 37712, and RestoreSensor, model 37714, Multi-program Rechargeable Neurostimulator for Spinal Cord Stimulation by Medtronic Neuromodulation, citing medtronic notified customers regarding Loss of Stimulation of specific Implantable Neuro-Stimulators under the following specific conditions: Switching between groups with Multipl….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-0278-2014
ClassificationClass II
Statusterminated
Initiated2013-09-30
Published2013-11-20
Last updated2026-09-13 11:53 UTC
CompanyMedtronic Neuromodulation
DistributionUS

Product

Medtronic RestoreUltra, model 37712, and RestoreSensor, model 37714, Multi-program Rechargeable Neurostimulator for Spinal Cord Stimulation.

Reason

Medtronic notified customers regarding Loss of Stimulation of specific Implantable Neuro-Stimulators under the following specific conditions: Switching between groups with Multiple programs or creating program groups. Therapy can be restored with the Physician Recharge mode (RPM) of the Implantable Neurostimulator Recharger (INSR).

Identifiers

code_infoall serial numbers

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:53 UTC. Current status: terminated.

Source

Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-0278-2014%22