FDA Device recall Z-0272-2016

Easy Dial Oxygen Regulator, 168708D and 168715D. Eight liter from .0.25 to 8 liters, 15 liter from 0.5 liter to 15 liters of oxygen per min…

On 2015-11-25 the FDA classified a Class II recall of Easy Dial Oxygen Regulator, 168708D and 168715D. Eight liter from .0.25 to 8 liters, 15 liter from 0.5 liter to 15 liters of oxygen per min… by Precision Medical, citing these lots of Oxygen Dial Regulators could have the settings label off by one position. When the indicator arrow points to OFF position, the regulator is actually on the half lite….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-0272-2016
ClassificationClass II
Statusterminated
Initiated2015-10-08
Published2015-11-25
Last updated2026-09-13 11:53 UTC
CompanyPrecision Medical
DistributionAL, AZ, CA, GA, ID, IN, KY, LA, MA, MI, MN, MO, MT, NC, ND, NY, OH, OK, PA, TX

Product

Easy Dial Oxygen Regulator, 168708D and 168715D. Eight liter from .0.25 to 8 liters, 15 liter from 0.5 liter to 15 liters of oxygen per minute.

Reason

These lots of Oxygen Dial Regulators could have the settings label off by one position. When the indicator arrow points to OFF position, the regulator is actually on the half liter setting (0.50), for the 15 liter. For the 8 liter, instead of being in the OFF position, the regulator will be in the quarter liter setting (0.25).

Identifiers

code_infoLot numbers 072215 and 072315.