FDA Device recall Z-0271-2019
O-arm 1000 2nd Edition Imaging System, OARM ASSY Bl70000027 SYSTEM PRODUCT, Model Number Bl70000027 Product Usage: The O-arm Imaging System…
- FDA Device
- Class II
- terminated
- Published 2018-10-31
On 2018-10-31 the FDA classified a Class II recall of O-arm 1000 2nd Edition Imaging System, OARM ASSY Bl70000027 SYSTEM PRODUCT, Model Number Bl70000027 Product Usage: The O-arm Imaging System… by Medtronic Navigation Littleton, citing software anomalies were reported to affect the following system functionalities: startup and shutdown, system and network communication, motion control, image acquisition and outp….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-0271-2019 |
| Classification | Class II |
| Status | terminated |
| Initiated | 2018-08-30 |
| Published | 2018-10-31 |
| Last updated | 2026-09-13 11:54 UTC |
| Company | Medtronic Navigation Littleton |
| Distribution | AK, AL, AR, AZ, CA, CO, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NY, OH, OK, OR, PA, PR, SC, SD, TN, TX, UT, VA, WA, WI |
Product
O-arm 1000 2nd Edition Imaging System, OARM ASSY Bl70000027 SYSTEM PRODUCT, Model Number Bl70000027 Product Usage: The O-arm Imaging System is a mobile x-ray system designed for 2D fluoroscopic and 3D imaging and is intended to be used where a physician benefits from 2D and 3D information of anatomic structures and objects with high x-ray attenuation such as bony anatomy and metallic objects. The 0-arm Imaging System is compatible with certain Image Guided Surgery Systems.
Reason
Software anomalies were reported to affect the following system functionalities: startup and shutdown, system and network communication, motion control, image acquisition and output, dose reporting, and logs generation. Software version 3.1.7 will be installed on all 2nd edition imaging systems.
Identifiers
| code_info | All Serial Numbers |
|---|