FDA Device recall Z-0270-2013
AC Powered adjustable hospital bed, Prime Bed TL500, Distributed by: Primus Medical LLC, 8401 Southern Blvd, Boardman, OH 44512
- FDA Device
- Class II
- terminated
- Published 2012-11-14
On 2012-11-14 the FDA classified a Class II recall of AC Powered adjustable hospital bed, Prime Bed TL500, Distributed by: Primus Medical LLC, 8401 Southern Blvd, Boardman, OH 44512 by Primus Medical, citing the "Retaining washers & Clevis Pins" provided with the AC -powered adjustable hospital beds may fail under certain circumstances. The pins at the top of the leg were disengaging….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-0270-2013 |
| Classification | Class II |
| Status | terminated |
| Initiated | 2011-02-02 |
| Published | 2012-11-14 |
| Last updated | 2026-09-13 11:53 UTC |
| Company | Primus Medical |
| Distribution | US |
Product
AC Powered adjustable hospital bed, Prime Bed TL500, Distributed by: Primus Medical LLC, 8401 Southern Blvd, Boardman, OH 44512. Usage: Hospital Bed.
Reason
The "Retaining washers & Clevis Pins" provided with the AC -powered adjustable hospital beds may fail under certain circumstances. The pins at the top of the leg were disengaging from the channel. This may cause the bed frame to come off tracks or collapse.
Identifiers
| code_info | Model #TL500; Serial Number: 001, 002, 003, 004, 005, 006, 007, 010, 013, 014, 015, 016, 017, 018, 021, 022, 023, 024,…Model #TL500; Serial Number: 001, 002, 003, 004, 005, 006, 007, 010, 013, 014, 015, 016, 017, 018, 021, 022, 023, 024, 025, 026, 027, 028, 029, 030, 031, 032, 033, 034, 035, 039, 040, 041, 042, 043, 044, 045, 046, 047, 048, 049, 050, 052, 053, 054, 055, 056, 057, 058, 059, 060, 061, 062, 063, 064, 065, 070, 075, 081, 09350014, 09350015, 09350016, 09350017, 09350018, 09350021, 09350023, 09350025, 09350026, 09350027, 09350028, 09350029, 09350030, 09350031, 09350032, 09350033, 09350035, 09350038, 09350040, 09350041, 09350042, 09350045, 09350048, 09350049, 09350054, 09350055, 09350056, 09350057, 09350058, 09350059, 09350060, 09350062, 09350063, 09350064, 09350065, 09350066, 09350067, 09350068, 09350069, 09350070, 09350072, 09350074, 09350082, 09350085, 09350087, 09350088, 09350089, 09350090, 09350091, 09350092, 09430001, 09430002, 09430003, 09430004, 09430005, 09430006, 09430007, 09430008, 09430010, 09430013, 09430014, 09430015, 09430017, 09430019, 09430020, 09430021, 09430022, 09430024, 0 |
|---|
Status history
No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:53 UTC. Current status: terminated.
Source
Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-0270-2013%22
RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.