FDA Device recall Z-0267-2022
Roberts Digital VIPR System (with E-size cylinder) Model: ROC-9840
- FDA Device
- Class II
- ongoing
- Published 2021-12-01
On 2021-12-01 the FDA classified a Class II recall of Roberts Digital VIPR System (with E-size cylinder) Model: ROC-9840 by Western Scott Fetzer, citing VIPR (Valve Integrated Pressure Regulator) - battery rupture, causing the plastic VIPR shroud to break apart with emission of fumes and smoke.
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-0267-2022 |
| Classification | Class II |
| Status | ongoing |
| Initiated | 2021-10-16 |
| Published | 2021-12-01 |
| Last updated | 2026-09-13 11:56 UTC |
| Company | Western Scott Fetzer |
| Distribution | CT, MD, MT, OH |
Product
Roberts Digital VIPR System (with E-size cylinder) Model: ROC-9840
Reason
VIPR (Valve Integrated Pressure Regulator) - battery rupture, causing the plastic VIPR shroud to break apart with emission of fumes and smoke
Identifiers
| code_info | Units distributed from 05/29/20 to 08/31/21. Digital VIPR units manufactured and distributed after 08/31/21 are outsiā¦Units distributed from 05/29/20 to 08/31/21. Digital VIPR units manufactured and distributed after 08/31/21 are outside the scope of this recall and will be recognized by a green and silver control knob sticker (green logo, silver background) |
|---|
Status history
No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:56 UTC. Current status: ongoing.
Source
Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-0267-2022%22
RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.