RecallRadar

FDA Device recall Z-0266-2026

Model Number SM-40HF-B-D-C

On 2025-10-29 the FDA classified a Class II recall of Model Number SM-40HF-B-D-C by Sedecal, citing sedecal sent an email to its consignees to remind the users that the equipment is not water-resistant and must be cleaned strictly according to the instructions in the manual, cle….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-0266-2026
ClassificationClass II
Statusongoing
Initiated2025-06-10
Published2025-10-29
Last updated2026-09-13 11:57 UTC
CompanySedecal
DistributionCA, IL, NJ

Product

Model Number SM-40HF-B-D-C; 40KW 55C/60C, Mobile X-ray system

Reason

Sedecal sent an email to its consignees to remind the users that the equipment is not water-resistant and must be cleaned strictly according to the instructions in the manual, clearly outlining the possible consequences of improper cleaning.

Identifiers

code_infoUDI/DI 08436046001510, Serial Numbers: G33585

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:57 UTC. Current status: ongoing.

Source

Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-0266-2026%22

RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.