Smith and Nephew EVOS Self-Tapping Locking Screw SST: (1)REF 72413575, 3.5MM x 75MM, and (2) REF 72413570, 3.5MM X 70MM. Sterile

On 2021-11-24 the FDA classified a Class II recall of Smith and Nephew EVOS Self-Tapping Locking Screw SST: (1)REF 72413575, 3.5MM x 75MM, and (2) REF 72413570, 3.5MM X 70MM. Sterile by Smith And Nephew, citing the labels on the packages of screws were switched.

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-0265-2022
ClassificationClass II
Statusongoing
Initiated2021-10-25
Published2021-11-24
Last updated2026-09-13 11:56 UTC
CompanySmith And Nephew
DistributionAL, AZ, CA, CO, CT, FL, GA, IL, IN, KY, LA, MA, ME, MI, MN, MO, NC, NJ, NY, OH, OR, PA, TN, TX, WA

Product

Smith and Nephew EVOS Self-Tapping Locking Screw SST: (1)REF 72413575, 3.5MM x 75MM, and (2) REF 72413570, 3.5MM X 70MM. Sterile.

Reason

The labels on the packages of screws were switched.

Identifiers

code_infoREF 72413575 - batch #21FM24289; REF 72413570 - batch #21GM04743.