FDA Device recall Z-0262-2022
Quantum 2000 Electrosurgical Generators Cooper Surgical Part Number: 909075-05
- FDA Device
- Class II
- ongoing
- Published 2021-11-24
On 2021-11-24 the FDA classified a Class II recall of Quantum 2000 Electrosurgical Generators Cooper Surgical Part Number: 909075-05 by Coopersurgical, citing coagulate function may not operate with the use of the Hand Switch, failure of the device to coagulate tissue may lead to a prolonged procedure requiring medical intervention to c….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-0262-2022 |
| Classification | Class II |
| Status | ongoing |
| Initiated | 2021-09-29 |
| Published | 2021-11-24 |
| Last updated | 2026-09-13 11:56 UTC |
| Company | Coopersurgical |
| Distribution | US |
Product
Quantum 2000 Electrosurgical Generators Cooper Surgical Part Number: 909075-05
Reason
Coagulate function may not operate with the use of the Hand Switch, failure of the device to coagulate tissue may lead to a prolonged procedure requiring medical intervention to control any intraoperative bleeding and complete the case.
Identifiers
| code_info | Serial Numbers: 0209001 0209002 0209003 0209004 0211001 0211002 0211003 0211004 0211005 0211006 0211007…Serial Numbers: 0209001 0209002 0209003 0209004 0211001 0211002 0211003 0211004 0211005 0211006 0211007 0211008 0211009 0211010 0211011 0211012 0211013 0211014 0211015 0211016 0211017 0211018 0211019 0211020 0211021 0211022 0211023 0211024 0211025 0211026 0211027 0211028 0211029 0211030 0211031 0211032 0211033 0211034 0211035 0211036 0211037 0211038 0211039 0211040 0215001 0215002 0215003 0215004 0215005 0215006 0215007 0215008 0215009 0215010 0215011 0215012 0215013 0215014 0215015 0215016 0215017 0215018 0215019 0215020 0215021 0215022 0215023 0215024 0215025 0215026 0215027 0215028 0215029 0215030 0215031 0215032 0215033 0215034 0215035 0215036 0314014 0314014KF 0314019 0410001 0410002 0410003 0410004 0410005 0410006 0410007 0410008 0410009 0410010 0410011 0410012 0410013 0410014 0410015 0410016 0410017 0410018 0410019 0410020 0410021 0410022 0410023 0410024 0410025 041002 |
|---|
Status history
No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:56 UTC. Current status: ongoing.
Source
Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-0262-2022%22