RecallRadar

FDA Device recall Z-0262-2018

Rezum System, Vapor BPH Ablation System, Model Number D2201-003

On 2017-12-27 the FDA classified a Class II recall of Rezum System, Vapor BPH Ablation System, Model Number D2201-003 by Nxthera, citing needle Bond may not be able to withstand the tension force applied when the shaped needle is deployed into and retracted from prostate tissue to deliver treatments.

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-0262-2018
ClassificationClass II
Statusterminated
Initiated2017-07-28
Published2017-12-27
Last updated2026-09-13 11:54 UTC
CompanyNxthera
DistributionAL, AZ, CA, CO, FL, GA, IL, MA, MI, MN, MO, NE, NJ, NY, OH, PA, TN, UT, WA

Product

Rezum System, Vapor BPH Ablation System, Model Number D2201-003. NxThera Inc. 7351 Kirkwood Lane Suite 138 Maple Grove, MN 55369 USA. Intended to relieve symptoms, obstructions, and reduce prostate tissue associated with benign prostatic hyperplasia (BPH). It is indicated for men greater or equal to 50 years of age with a prostate volume greater than or equal to 30cm3 and less than or equal to 80cm3. The Rezum System is also indicated for treatment of prostates with hyperplasia of the central zone and/or a median lobe.

Reason

Needle Bond may not be able to withstand the tension force applied when the shaped needle is deployed into and retracted from prostate tissue to deliver treatments.

Identifiers

code_info
UDI 08855757006010. Lot Number of Outside Cases (Pack Lot) 2017020920 with Lot Number of Inside Boxes (Product Lot) 201…UDI 08855757006010. Lot Number of Outside Cases (Pack Lot) 2017020920 with Lot Number of Inside Boxes (Product Lot) 2017011612 and Exp Date 2018-01-31: Lot Number of Outside Cases (Pack Lot) 2017030286 with Lot Number of Inside Boxees (Product Lot) 2017011187 and Exp Date 2018-01-31.

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:54 UTC. Current status: terminated.

Source

Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-0262-2018%22

RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.