FDA Device recall Z-0261-2020
Implant Direct Scan Adapter Assembly Replant platform for the following part numbers: 6035-09PT-M (3.5mmD)/UDI : 10841307123934,;6043-09PT-…
- FDA Device
- Class II
- terminated
- Published 2019-11-06
On 2019-11-06 the FDA classified a Class II recall of Implant Direct Scan Adapter Assembly Replant platform for the following part numbers: 6035-09PT-M (3.5mmD)/UDI : 10841307123934,;6043-09PT-… by Implant Direct Sybron Manufacturing, citing the action is being initiated due to incorrect assembly of the scan adapter . The result of the incorrect assembly could cause the clinician not to be able to capture the position….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-0261-2020 |
| Classification | Class II |
| Status | terminated |
| Initiated | 2018-09-06 |
| Published | 2019-11-06 |
| Last updated | 2026-09-13 11:55 UTC |
| Company | Implant Direct Sybron Manufacturing |
| Distribution | CA, CO, FL, GA, LA, MA, NE, NJ, NY, OH, PA, SC, UT, VA |
Product
Implant Direct Scan Adapter Assembly Replant platform for the following part numbers: 6035-09PT-M (3.5mmD)/UDI : 10841307123934,;6043-09PT-M (4.3mmD)/UDI : 10841307123941; and 6050-09PT-M (5.00mmD)/ UDI: 10841307123958
Reason
The action is being initiated due to incorrect assembly of the scan adapter . The result of the incorrect assembly could cause the clinician not to be able to capture the position and orientation of a dental implant(s) or lab analog during the digital scanning process which could result in a delay for the patient.
Identifiers
| code_info | Part numbers/Lot Numbers: 6035-09PT/Lot # 104539; 6043-09PT/ Lot # 104540; 6050-09PT/Lot # 104511 |
|---|
Status history
No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:55 UTC. Current status: terminated.
Source
Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-0261-2020%22