RecallRadar

FDA Device recall Z-0259-2021

5.5" (14 cm) Appx 0.39 ml, Smallbore Trifuse Ext Set w/3 MicroClave Clear, NanoClave T-Connector, 3 Clamps, Rotating Luer.1 unit per pouch,…

On 2020-10-28 the FDA classified a Class II recall of 5.5" (14 cm) Appx 0.39 ml, Smallbore Trifuse Ext Set w/3 MicroClave Clear, NanoClave T-Connector, 3 Clamps, Rotating Luer.1 unit per pouch,… by Icu Medical, citing identification of a potential manufacturing defect on the internal surface of the NanoClave within specific lots of NanoClave sets, which may inhibit a proper seal with the NanoCl….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-0259-2021
ClassificationClass II
Statusterminated
Initiated2020-07-06
Published2020-10-28
Last updated2026-09-13 11:55 UTC
CompanyIcu Medical
DistributionUS

Product

5.5" (14 cm) Appx 0.39 ml, Smallbore Trifuse Ext Set w/3 MicroClave Clear, NanoClave T-Connector, 3 Clamps, Rotating Luer.1 unit per pouch, 50 multivac pouches per case. UDI: (01)10887709032153(17)241101(30)50(10)4475297; (01)10887709032153(17)250201(30)50(10)4711598. The Intravascular Administration Set is a single use, sterile, non-pyrogenic device which provides access for the administration of fluids from a container to a patient's vascular system through the administration set's needle or catheter (which is inserted into a vein).

Reason

Identification of a potential manufacturing defect on the internal surface of the NanoClave within specific lots of NanoClave sets, which may inhibit a proper seal with the NanoClave spike.

Identifiers

code_infoLot Numbers:4475297, 4711598.

Status history