RecallRadar

FDA Device recall Z-0259-2013

AMS AdVance" Male Sling System 720088-01 AMS AdVance" Male Sling System (US) 720088-02 AMS AdVance" Male Sling System (Intl) 72404224 AMS M…

On 2012-11-14 the FDA classified a Class II recall of AMS AdVance" Male Sling System 720088-01 AMS AdVance" Male Sling System (US) 720088-02 AMS AdVance" Male Sling System (Intl) 72404224 AMS M… by American Medical Systems, citing american Medical Systems¿ (AMS) is issuing a voluntary recall on a number of lots of the AdVance¿ Male Sling System (AdVance). AMS has recently received a small number of complain….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-0259-2013
ClassificationClass II
Statusterminated
Initiated2012-10-19
Published2012-11-14
Last updated2026-09-13 11:53 UTC
CompanyAmerican Medical Systems
DistributionUS

Product

AMS AdVance" Male Sling System 720088-01 AMS AdVance" Male Sling System (US) 720088-02 AMS AdVance" Male Sling System (Intl) 72404224 AMS Male TO Sling System (Intl). The AMS AdVance Male Sling System is intended for the placement of a suburethral sling for the treatment of male stress urinary incontinence (SUI).

Reason

American Medical Systems¿ (AMS) is issuing a voluntary recall on a number of lots of the AdVance¿ Male Sling System (AdVance). AMS has recently received a small number of complaints regarding the resorbable suture in the AdVance sling not lasting for the entire 15 month shelf life of the product. An internal investigation has been conducted and confirmed that some sutures degraded prior to the

Identifiers

code_info
Part # Lot/Serial Range 720088-01 722022001 to 796127002 720088-02 722024001 to 7…Part # Lot/Serial Range 720088-01 722022001 to 796127002 720088-02 722024001 to 790266012 72404224 742528001 to 795689010

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:53 UTC. Current status: terminated.

Source

Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-0259-2013%22

RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.