FDA Device recall Z-0259-2013
AMS AdVance" Male Sling System 720088-01 AMS AdVance" Male Sling System (US) 720088-02 AMS AdVance" Male Sling System (Intl) 72404224 AMS M…
- FDA Device
- Class II
- terminated
- Published 2012-11-14
On 2012-11-14 the FDA classified a Class II recall of AMS AdVance" Male Sling System 720088-01 AMS AdVance" Male Sling System (US) 720088-02 AMS AdVance" Male Sling System (Intl) 72404224 AMS M… by American Medical Systems, citing american Medical Systems¿ (AMS) is issuing a voluntary recall on a number of lots of the AdVance¿ Male Sling System (AdVance). AMS has recently received a small number of complain….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-0259-2013 |
| Classification | Class II |
| Status | terminated |
| Initiated | 2012-10-19 |
| Published | 2012-11-14 |
| Last updated | 2026-09-13 11:53 UTC |
| Company | American Medical Systems |
| Distribution | US |
Product
AMS AdVance" Male Sling System 720088-01 AMS AdVance" Male Sling System (US) 720088-02 AMS AdVance" Male Sling System (Intl) 72404224 AMS Male TO Sling System (Intl). The AMS AdVance Male Sling System is intended for the placement of a suburethral sling for the treatment of male stress urinary incontinence (SUI).
Reason
American Medical Systems¿ (AMS) is issuing a voluntary recall on a number of lots of the AdVance¿ Male Sling System (AdVance). AMS has recently received a small number of complaints regarding the resorbable suture in the AdVance sling not lasting for the entire 15 month shelf life of the product. An internal investigation has been conducted and confirmed that some sutures degraded prior to the
Identifiers
| code_info | Part # Lot/Serial Range 720088-01 722022001 to 796127002 720088-02 722024001 to 7…Part # Lot/Serial Range 720088-01 722022001 to 796127002 720088-02 722024001 to 790266012 72404224 742528001 to 795689010 |
|---|
Status history
No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:53 UTC. Current status: terminated.
Source
Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-0259-2013%22