RecallRadar

FDA Device recall Z-0245-2023

Infection Control (IC) Lifting Strap used with the following Cirrus and Pinnacle Ceiling Lifts: Cirrus 450, Cirrus 600, Cirrus 700, Cirrus…

On 2022-11-23 the FDA classified a Class II recall of Infection Control (IC) Lifting Strap used with the following Cirrus and Pinnacle Ceiling Lifts: Cirrus 450, Cirrus 600, Cirrus 700, Cirrus… by Tollos, citing two customers reported four incidents in which the lifting strap broke while lifting a patient. If the IC strap fails during use, the patient may fall causing injury to them or th….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-0245-2023
ClassificationClass II
Statusongoing
Initiated2022-10-24
Published2022-11-23
Last updated2026-09-13 11:56 UTC
CompanyTollos
DistributionUS

Product

Infection Control (IC) Lifting Strap used with the following Cirrus and Pinnacle Ceiling Lifts: Cirrus 450, Cirrus 600, Cirrus 700, Cirrus 750, and Pinnacle 1000

Reason

Two customers reported four incidents in which the lifting strap broke while lifting a patient. If the IC strap fails during use, the patient may fall causing injury to them or the caregiver(s).

Identifiers

code_infoUDI-DI 00817956021 (not yet uploaded to GUDID)

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:56 UTC. Current status: ongoing.

Source

Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-0245-2023%22

RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.