Computed Tomography Systems Revolution CT, Revolution CT ES, Revolution Apex, Revolution CT with Apex edition

On 2021-12-01 the FDA classified a Class II recall of Computed Tomography Systems Revolution CT, Revolution CT ES, Revolution Apex, Revolution CT with Apex edition by Ge Healthcare, citing the accumulated dose is incorrectly displayed in certain situations. Thisoccurs during an interventional exam using the SmartStep option if the table height is adjusted after the….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-0242-2022
ClassificationClass II
Statusongoing
Initiated2021-06-30
Published2021-12-01
Last updated2026-09-13 11:56 UTC
CompanyGe Healthcare
DistributionUS

Product

Computed Tomography Systems Revolution CT, Revolution CT ES, Revolution Apex, Revolution CT with Apex edition

Reason

The accumulated dose is incorrectly displayed in certain situations. Thisoccurs during an interventional exam using the SmartStep option if the table height is adjusted after the exam starts. In this situation, the SmartStep display will show a value that is higher (up to 100 times) than the actual dose delivered.

Identifiers

code_infoRevolution CT, Revolution CT ES, Revolution Apex, Revolution CT with Apex edition