FDA Device recall Z-0241-2023
GI4000 Electrosurgical Unit - intended to deliver electrosurgical outputs to perform cutting and/or coagulation and coagulation with or wit…
- FDA Device
- Class II
- terminated
- Published 2022-11-23
On 2022-11-23 the FDA classified a Class II recall of GI4000 Electrosurgical Unit - intended to deliver electrosurgical outputs to perform cutting and/or coagulation and coagulation with or wit… by Steris, citing two electronic diode components (D1 & D4) placed in the reverse orientation on the controller printed circuit board assembly may result in premature wear of the power regulator, l….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-0241-2023 |
| Classification | Class II |
| Status | terminated |
| Initiated | 2022-10-17 |
| Published | 2022-11-23 |
| Last updated | 2026-09-13 11:56 UTC |
| Company | Steris |
| Distribution | US |
Product
GI4000 Electrosurgical Unit - intended to deliver electrosurgical outputs to perform cutting and/or coagulation and coagulation with or without delivery of argon gas, in flexible endoscopic applications. MODEL #(s): G1110001
Reason
Two electronic diode components (D1 & D4) placed in the reverse orientation on the controller printed circuit board assembly may result in premature wear of the power regulator, leading to the regulators failure which could cause a full power loss. If the regulator failure were to occur during a patient procedure, a procedural delay may result
Identifiers
| code_info | UDI: 00724995180256 Serial Numbers: 0032-0822-003, 0032-0622-009, 0032-0922-001, 0032-0822-001, and 0032-0082-002 |
|---|