RecallRadar

FDA Device recall Z-0241-2023

GI4000 Electrosurgical Unit - intended to deliver electrosurgical outputs to perform cutting and/or coagulation and coagulation with or wit…

On 2022-11-23 the FDA classified a Class II recall of GI4000 Electrosurgical Unit - intended to deliver electrosurgical outputs to perform cutting and/or coagulation and coagulation with or wit… by Steris, citing two electronic diode components (D1 & D4) placed in the reverse orientation on the controller printed circuit board assembly may result in premature wear of the power regulator, l….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-0241-2023
ClassificationClass II
Statusterminated
Initiated2022-10-17
Published2022-11-23
Last updated2026-09-13 11:56 UTC
CompanySteris
DistributionUS

Product

GI4000 Electrosurgical Unit - intended to deliver electrosurgical outputs to perform cutting and/or coagulation and coagulation with or without delivery of argon gas, in flexible endoscopic applications. MODEL #(s): G1110001

Reason

Two electronic diode components (D1 & D4) placed in the reverse orientation on the controller printed circuit board assembly may result in premature wear of the power regulator, leading to the regulators failure which could cause a full power loss. If the regulator failure were to occur during a patient procedure, a procedural delay may result

Identifiers

code_infoUDI: 00724995180256 Serial Numbers: 0032-0822-003, 0032-0622-009, 0032-0922-001, 0032-0822-001, and 0032-0082-002

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:56 UTC. Current status: terminated.

Source

Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-0241-2023%22

RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.