Hill-Rom Progressa Bed System

On 2022-11-23 the FDA classified a Class II recall of Hill-Rom Progressa Bed System by Baxter Healthcare, citing the compression links on impacted Progressa beds, listed in Appendix A, may bend when raising the head section. The labeling for these beds list a maximum patient weight of 500 lb….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-0231-2023
ClassificationClass II
Statusongoing
Initiated2022-10-18
Published2022-11-23
Last updated2026-09-13 11:56 UTC
CompanyBaxter Healthcare
DistributionUS

Product

Hill-Rom Progressa Bed System

Reason

The compression links on impacted Progressa beds, listed in Appendix A, may bend when raising the head section. The labeling for these beds list a maximum patient weight of 500 lbs., however the bending may still occur with a patient weight of less than 500 lbs. depending on the patient 's position on the bed.

Identifiers

code_info
UDI/DI 00887761000100, Serial Numbers: X157AW7312, X206AW8017, X154AW7292, X166AW7416, X207AW8041, X158AW7351, X157AW7…UDI/DI 00887761000100, Serial Numbers: X157AW7312, X206AW8017, X154AW7292, X166AW7416, X207AW8041, X158AW7351, X157AW7326, X157AW7313, X206AW8018, X154AW7293, X166AW7418, X167AW7449, X158AW7352, X157AW7328, X174AW7575, X206AW8019, X154AW7296, X167AW7428, X167AW7450, X158AW7353, X157AW7329, X174AW7579, X206AW8021, X154AW7297, X167AW7429, X167AW7451, X158AW7354, X157AW7330, X181AW7722, X206AW8022, X154AW7298, X157AW7308, X166AW7419, X158AW7355, X157AW7332, X181AW7728, X206AW8023, X154AW7300, X157AW7309, X167AW7425, X206AW8010, X158AW7340, X182AW7736, X206AW8024, X168AW7456, X178AW7622, X167AW7427, X206AW8011, X158AW7341, X182AW7737, X206AW8025, X168AW7478, X178AW7623, X159AW7375, X206AW8012, X158AW7342, X207AW8060, X158AW7344, X171AW7495, X178AW7625, X159AW7378, X206AW8020, X158AW7343, X208AW8079, X206AW8026, X171AW7496, X178AW7626, X159AW7380, X158AW7333, X158AW7349, X208AW8082, X206AW8027, X171AW7497, X178AW7627, X216AW8193, X158AW7334, X158AW7359, X208AW8084, X206AW8028, X171AW7498,

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:56 UTC. Current status: ongoing.

Source

Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-0231-2023%22

RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.