RecallRadar

FDA Device recall Z-0230-2023

Superion¿ Indirect Decompression System (IDS) Model Numbers and Size: 101-9808 - 8mm 01-9812 - 12mm 101-9814 - 14mm

On 2022-11-23 the FDA classified a Class II recall of Superion¿ Indirect Decompression System (IDS) Model Numbers and Size: 101-9808 - 8mm 01-9812 - 12mm 101-9814 - 14mm by Boston Scientific Neuromodulation, citing due to Indirect Decompression System devices exhibiting a potential for weld breakage at the spindle cap-to-main body interface. As a result of the separation of the spindle cap f….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-0230-2023
ClassificationClass II
Statusongoing
Initiated2022-10-12
Published2022-11-23
Last updated2026-09-13 11:56 UTC
CompanyBoston Scientific Neuromodulation
DistributionUS

Product

Superion¿ Indirect Decompression System (IDS) Model Numbers and Size: 101-9808 - 8mm 01-9812 - 12mm 101-9814 - 14mm

Reason

Due to Indirect Decompression System devices exhibiting a potential for weld breakage at the spindle cap-to-main body interface. As a result of the separation of the spindle cap from the main body it may prevent proper deployment of the device during implantation.

Identifiers

code_info
Model (UPN): 101-9808 UDI-DI (GTIN) Code: 00884662000529 Lot/Batch Numbers: 26176977 Model (UPN): 101-9812 UDI-DI (GTI…Model (UPN): 101-9808 UDI-DI (GTIN) Code: 00884662000529 Lot/Batch Numbers: 26176977 Model (UPN): 101-9812 UDI-DI (GTIN) Code: 00884662000543 Lot/Batch Numbers: 26573432 26375758 26815560 26807662 26654632 26928168 26929605 26489700 Model (UPN): 101-9814 UDI-DI (GTIN) Code: 00884662000550 Lot/Batch Numbers: 26366842 26848047 26619510 26323984 26406210 26799951

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:56 UTC. Current status: ongoing.

Source

Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-0230-2023%22

RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.