FDA Device recall Z-0229-2022
bvi CustomEyes Kit-Lafayette General Medical Center: CustomEyes Cataract Surgical Kits that contain Beaver Xstar Crescent Knife (REF#373809…
- FDA Device
- Class II
- ongoing
- Published 2021-11-24
On 2021-11-24 the FDA classified a Class II recall of bvi CustomEyes Kit-Lafayette General Medical Center: CustomEyes Cataract Surgical Kits that contain Beaver Xstar Crescent Knife (REF#373809… by Beaver Visitec International, citing product packaged as Beaver Xstar Crescent Knife, 2.5mm, 55 degrees bevel-down blades contain bevel-up blades.
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-0229-2022 |
| Classification | Class II |
| Status | ongoing |
| Initiated | 2021-10-08 |
| Published | 2021-11-24 |
| Last updated | 2026-09-13 11:56 UTC |
| Company | Beaver Visitec International |
| Distribution | US |
Product
bvi CustomEyes Kit-Lafayette General Medical Center: CustomEyes Cataract Surgical Kits that contain Beaver Xstar Crescent Knife (REF#373809) as a component REF# 58000572 REF#: 373809
Reason
Product packaged as Beaver Xstar Crescent Knife, 2.5mm, 55 degrees bevel-down blades contain bevel-up blades
Identifiers
| code_info | Lot Numbers 6032947 and 6033588 UDI: " Primary: 00886158019961 " Secondary: 30886158019962 " Tertiary: 50886158019966 |
|---|