- Published 2022-11-16
On 2022-11-16 the FDA classified a Class II recall of ORISE Gel Syringe Submucosal Lifting Agent Kits: indicated for use in gastrointestinal endoscopic procedures for submucosal lift of polyps,… by Boston Scientific, citing ***Update12/13/22***:Product removal due to higher incidence of adverse events of foreign body reactions*** Potential outcomes associated with foreign body reaction to remnant ORl….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-0225-2023 |
| Classification | Class II |
| Status | ongoing |
| Initiated | 2022-10-10 |
| Published | 2022-11-16 |
| Last updated | 2026-09-13 11:56 UTC |
| Company | Boston Scientific |
| Distribution | US |
Product
ORISE Gel Syringe Submucosal Lifting Agent Kits: indicated for use in gastrointestinal endoscopic procedures for submucosal lift of polyps, adenomas, early-stage cancers or other gastrointestinal mucosal lesions, prior to excision with a snare or other suitable endoscopic device UPN: (1)Twin Pack Box 1 M00519210; (2)Twin Pack Box 10 M00519211; (3)Single Pack Box 10 M00519231
Reason
***Update12/13/22***:Product removal due to higher incidence of adverse events of foreign body reactions*** Potential outcomes associated with foreign body reaction to remnant ORlSE Gel post procedure. Foreign body reaction which presented as mass formations from remnant ORlSE Gel post-procedure, prompting surgical intervention
Identifiers
| code_info | GTIN Number : (1)08714729974581 (2)08714729974598 (3)08714729993865 All batches |
|---|
Status history
No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:56 UTC. Current status: ongoing.
Source
Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-0225-2023%22
RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.