RecallRadar

FDA Device recall Z-0224-2026

Shanghai United Imaging Healthcare Co., Ltd

On 2025-10-29 the FDA classified a Class II recall of Shanghai United Imaging Healthcare Co., Ltd by Uih Technologies, citing third-party outer covers can become trapped between patient table and gantry, causing mechanical interference during table movement.

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-0224-2026
ClassificationClass II
Statusongoing
Initiated2025-09-12
Published2025-10-29
Last updated2026-09-13 11:57 UTC
CompanyUih Technologies
Distributionn/a

Product

Shanghai United Imaging Healthcare Co., Ltd. Positron Emission Tomography and Computed Tomography System Model:uMI 780 MD, Rx Only

Reason

Third-party outer covers can become trapped between patient table and gantry, causing mechanical interference during table movement.

Identifiers

code_infoUDI: (01)06971576832033/ Serial Number: 300072, 300082, 300098

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:57 UTC. Current status: ongoing.

Source

Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-0224-2026%22

RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.