FDA Device recall Z-0224-2021

3-Port NanoClave Manifold w/Check Valve, Rotating Luer. 1 unit per pouch, 50 pouches per case

On 2020-10-28 the FDA classified a Class II recall of 3-Port NanoClave Manifold w/Check Valve, Rotating Luer. 1 unit per pouch, 50 pouches per case by Icu Medical, citing identification of a potential manufacturing defect on the internal surface of the NanoClave within specific lots of NanoClave sets, which may inhibit a proper seal with the NanoCl….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-0224-2021
ClassificationClass II
Statusterminated
Initiated2020-07-06
Published2020-10-28
Last updated2026-09-13 11:55 UTC
CompanyIcu Medical
DistributionUS

Product

3-Port NanoClave Manifold w/Check Valve, Rotating Luer. 1 unit per pouch, 50 pouches per case. UDI:(01)10887709087153(17)250101(30)50(10)4556715; (01)10887709087153(17)250101(30)50(10)4579939; (01)10887709087153(17)250101(30)50(10)4592179; (01)10887709087153(17)250101(30)50(10)4610925; (01)10887709087153(17)241201(30)50(10)4547212; (01)10887709087153(17)250101(30)50(10)4712946; (01)10887709087153(17)250201(30)50(10)4720392. The Intravascular Administration Set is a single use, sterile, non-pyrogenic device which provides access for the administration of fluids from a container to a patient's vascular system through the administration set's needle or catheter (which is inserted into a vein).

Reason

Identification of a potential manufacturing defect on the internal surface of the NanoClave within specific lots of NanoClave sets, which may inhibit a proper seal with the NanoClave spike.

Identifiers

code_infoLot Numbers:4556715, 4579939, 4592179, 4610925, 4547212, 4712946,4720392.

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:55 UTC. Current status: terminated.