LeadCare II Blood Lead Test Kit-For in vitro diagnostic use only

On 2023-11-15 the FDA classified a Class II recall of LeadCare II Blood Lead Test Kit-For in vitro diagnostic use only by Magellan Diagnostics, citing during an internal audit a small percentage (<10%) of plastic caps used to close the treatment reagent tubes were found to not meet the minimum dimensional lip requirement and may….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-0219-2024
ClassificationClass II
Statusongoing
Initiated2023-09-19
Published2023-11-15
Last updated2026-09-13 11:56 UTC
CompanyMagellan Diagnostics
DistributionUS

Product

LeadCare II Blood Lead Test Kit-For in vitro diagnostic use only. The test kit is for the quantitative measurement of lead in capillary whole blood.

Reason

During an internal audit a small percentage (<10%) of plastic caps used to close the treatment reagent tubes were found to not meet the minimum dimensional lip requirement and may have inadequate cap-to-tube seals. Manufacturing with these caps may result in cap-to-tube seals that allow leaking of HCl or may result in cap-to-tube seals that allow evaporation of treatment reagent. The consequence of this would be a change in the volume of treatment reagent and a change in the ratios of either added patient sample or control material to treatment reagent. The result of this might be a falsely elevated result (e.g., a patient false positive [PATFP] result or a Controls Out of Range-High [COOR-Hi] result).

Identifiers

code_info
UDI-DI :(01)00850355006000 Lot Numbers: 2231 2231-10 2303M-11 2303M-12 2303M-13 2303M-15 2305M-01 2305M-02 2305M-03 23…UDI-DI :(01)00850355006000 Lot Numbers: 2231 2231-10 2303M-11 2303M-12 2303M-13 2303M-15 2305M-01 2305M-02 2305M-03 2305M-04 2305M-05 2306M-01 2306M-02 2306M-03 2306M-04 2306M-05 2306M-06 2307M 2307M-04 2307M-05 2307M-06 2307M-07 2307M-10 2307M-11 2308M-01 2308M-02 2308M-03 2308M-07 2309M 2310M-01 2310M-03 2310M-04 2310M-05 2310M-06 2310M-08 2311M-02 2311M-03 2311M-05 2311M-09 2311M-10 2312M 2312M-01 2312M-04 2312M-05 2312M-07 2313M 2313M-01 2313M-05 2313M-07 2314M-06 2314M-07 2314M-08 2314M-09 2314M-12 2314M-13 2315M-02 2315M-03 2315M-04 2315M-05 2315M-06 2315M-07 2315M-08 2315M-13 2316M 2316M-03 2316M-04 2316M-05 2316M-06 2316M-11 2316M-12 2317M 2317M-01 2318M 2320M-06 7308M