FDA Device recall Z-0219-2024
LeadCare II Blood Lead Test Kit-For in vitro diagnostic use only
- FDA Device
- Class II
- ongoing
- Published 2023-11-15
On 2023-11-15 the FDA classified a Class II recall of LeadCare II Blood Lead Test Kit-For in vitro diagnostic use only by Magellan Diagnostics, citing during an internal audit a small percentage (<10%) of plastic caps used to close the treatment reagent tubes were found to not meet the minimum dimensional lip requirement and may….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-0219-2024 |
| Classification | Class II |
| Status | ongoing |
| Initiated | 2023-09-19 |
| Published | 2023-11-15 |
| Last updated | 2026-09-13 11:56 UTC |
| Company | Magellan Diagnostics |
| Distribution | US |
Product
LeadCare II Blood Lead Test Kit-For in vitro diagnostic use only. The test kit is for the quantitative measurement of lead in capillary whole blood.
Reason
During an internal audit a small percentage (<10%) of plastic caps used to close the treatment reagent tubes were found to not meet the minimum dimensional lip requirement and may have inadequate cap-to-tube seals. Manufacturing with these caps may result in cap-to-tube seals that allow leaking of HCl or may result in cap-to-tube seals that allow evaporation of treatment reagent. The consequence of this would be a change in the volume of treatment reagent and a change in the ratios of either added patient sample or control material to treatment reagent. The result of this might be a falsely elevated result (e.g., a patient false positive [PATFP] result or a Controls Out of Range-High [COOR-Hi] result).
Identifiers
| code_info | UDI-DI :(01)00850355006000 Lot Numbers: 2231 2231-10 2303M-11 2303M-12 2303M-13 2303M-15 2305M-01 2305M-02 2305M-03 23…UDI-DI :(01)00850355006000 Lot Numbers: 2231 2231-10 2303M-11 2303M-12 2303M-13 2303M-15 2305M-01 2305M-02 2305M-03 2305M-04 2305M-05 2306M-01 2306M-02 2306M-03 2306M-04 2306M-05 2306M-06 2307M 2307M-04 2307M-05 2307M-06 2307M-07 2307M-10 2307M-11 2308M-01 2308M-02 2308M-03 2308M-07 2309M 2310M-01 2310M-03 2310M-04 2310M-05 2310M-06 2310M-08 2311M-02 2311M-03 2311M-05 2311M-09 2311M-10 2312M 2312M-01 2312M-04 2312M-05 2312M-07 2313M 2313M-01 2313M-05 2313M-07 2314M-06 2314M-07 2314M-08 2314M-09 2314M-12 2314M-13 2315M-02 2315M-03 2315M-04 2315M-05 2315M-06 2315M-07 2315M-08 2315M-13 2316M 2316M-03 2316M-04 2316M-05 2316M-06 2316M-11 2316M-12 2317M 2317M-01 2318M 2320M-06 7308M |
|---|