FDA Device recall Z-0219-2021

113" (287 cm) Appx 14.7 ml, 10 Drop Admin Set w/MicroClave" Clear, Remv 4-Way NanoClave" Stopcock, Remv 2-Gang 4-Way NanoClave" Stopcocks,…

On 2020-10-28 the FDA classified a Class II recall of 113" (287 cm) Appx 14.7 ml, 10 Drop Admin Set w/MicroClave" Clear, Remv 4-Way NanoClave" Stopcock, Remv 2-Gang 4-Way NanoClave" Stopcocks,… by Icu Medical, citing identification of a potential manufacturing defect on the internal surface of the NanoClave within specific lots of NanoClave sets, which may inhibit a proper seal with the NanoCl….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-0219-2021
ClassificationClass II
Statusterminated
Initiated2020-07-06
Published2020-10-28
Last updated2026-09-13 11:55 UTC
CompanyIcu Medical
DistributionUS

Product

113" (287 cm) Appx 14.7 ml, 10 Drop Admin Set w/MicroClave" Clear, Remv 4-Way NanoClave" Stopcock, Remv 2-Gang 4-Way NanoClave" Stopcocks, Spin Luer w/ Filter Cap, 2 Ext.1 unit per pouch, 50 multivac pouches per case. UDI:(01)10887709120171(17)250201(30)50(10)4734446. The Intravascular Administration Set is a single use, sterile, non-pyrogenic device which provides access for the administration of fluids from a container to a patient's vascular system through the administration set's needle or catheter (which is inserted into a vein).

Reason

Identification of a potential manufacturing defect on the internal surface of the NanoClave within specific lots of NanoClave sets, which may inhibit a proper seal with the NanoClave spike.

Identifiers

code_infoLot Number:4734446.

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:55 UTC. Current status: terminated.

Source

Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-0219-2021%22