FDA Device recall Z-0215-2026
DxC 500i Clinical Analyzer Modules, DxC 500 AU Module w/ISE, DxC 500i (Part Number C63522), DxC 500 AU Module, DxC 500i (Part Number C63521…
- FDA Device
- Class II
- ongoing
- Published 2025-10-29
On 2025-10-29 the FDA classified a Class II recall of DxC 500i Clinical Analyzer Modules, DxC 500 AU Module w/ISE, DxC 500i (Part Number C63522), DxC 500 AU Module, DxC 500i (Part Number C63521… by Beckman Coulter, citing beckman Coulter has confirmed a software anomaly on DxI 9000 Access Immunoassay Analyzer and the DxC 500i Clinical Analyzer that may prevent correct application of user-specified….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-0215-2026 |
| Classification | Class II |
| Status | ongoing |
| Initiated | 2025-09-19 |
| Published | 2025-10-29 |
| Last updated | 2026-09-13 11:57 UTC |
| Company | Beckman Coulter |
| Distribution | US |
Product
DxC 500i Clinical Analyzer Modules, DxC 500 AU Module w/ISE, DxC 500i (Part Number C63522), DxC 500 AU Module, DxC 500i (Part Number C63521), and Access 2 Module, DxC 500i (Part Number C13252)
Reason
Beckman Coulter has confirmed a software anomaly on DxI 9000 Access Immunoassay Analyzer and the DxC 500i Clinical Analyzer that may prevent correct application of user-specified manual dilution factors when ordering additional tests for samples already "In Progress" using the same "Sample ID number". This can lead to false low or high patient results and potential delays in reporting, risking erroneous diagnoses if undetected.
Identifiers
| code_info | C63522 UDI-DI: 14987666545089; C63521 UDI-DI: 14987666545072; C13252 UDI-DI: 15099590742331; Serial Numbers: 750198 75…C63522 UDI-DI: 14987666545089; C63521 UDI-DI: 14987666545072; C13252 UDI-DI: 15099590742331; Serial Numbers: 750198 750058 750222 750128 750138 750069 750050 750055 750179 750180 750181 750109 750168 750104 750211 750073 750210 750209 750178 750151 750153 750152 750155 750185 750187 750188 750195 750196 750201 750197 750199 750204 750202 750205 750083 750079 750186 750052 750230 750228 750140 750077 750135 750130 750167 750095 750105 750175 750132 750082 750113 750221 750219 750100 750099 750174 750053 750246 750134 750217 750216 750236 750233 750143 750081 750085 750071 750208 750207 750076 750131 750148 750223 750137 750116 750166 750241 750240 750183 750193 750200 750070 750087 750239 750107 750093 750124 750215 750141 750145 750144 750117 750108 750184 750173 750224 750218 750123 750129 750054 750206 750146 750101 750112 750194 750115 750157 750159 750160 750162 750125 750103 750229 750226 750234 750133 750048 750163 750182 750192 750189 750158 750250 750214 750110 750247 750154 7 |
|---|
Status history
No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:57 UTC. Current status: ongoing.
Source
Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-0215-2026%22