RecallRadar

FDA Device recall Z-0214-2016

Eon 100 system included the following product reference numbers and names: Reference no. 77345- Glucose (GPSL) Reference no. 77450- Uric ac…

On 2015-11-11 the FDA classified a Class II recall of Eon 100 system included the following product reference numbers and names: Reference no. 77345- Glucose (GPSL) Reference no. 77450- Uric ac… by Elitech Clinical Systems Sas, citing ELITech clinical Systems reagents based on Trinder reaction used on the Eon 100 instrument system are recalled due to potential drug interference. No complaints were received.

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-0214-2016
ClassificationClass II
Statusterminated
Initiated2015-09-22
Published2015-11-11
Last updated2026-09-13 11:53 UTC
CompanyElitech Clinical Systems Sas
DistributionUS

Product

Eon 100 system included the following product reference numbers and names: Reference no. 77345- Glucose (GPSL) Reference no. 77450- Uric acid (AUVD) Reference no. 77435- Triglycerides (TGML) Reference no. 77295- Cholesterol (CHVD) Reference no. 77301- Cholesterol HDL (HDLL) Reference no. 77310- Cholesterol, LDL (LDLL)

Reason

ELITech clinical Systems reagents based on Trinder reaction used on the Eon 100 instrument system are recalled due to potential drug interference. No complaints were received.