Fisher and Paykel Healthcare (FPH) IW900-Series Infant Warmer Fisher and Paykel

On 2014-11-19 the FDA classified a Class II recall of Fisher and Paykel Healthcare (FPH) IW900-Series Infant Warmer Fisher and Paykel by Fisher And Paykel Healthcare, citing breakage of the nut that secures the heater head in place, which can cause the heater head to become partially detached and swing towards the bassinet. There is the potential that….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-0210-2015
ClassificationClass II
Statusterminated
Initiated2014-11-06
Published2014-11-19
Last updated2026-09-13 11:53 UTC
CompanyFisher And Paykel Healthcare
DistributionUS

Product

Fisher and Paykel Healthcare (FPH) IW900-Series Infant Warmer Fisher and Paykel; Healthcare CosyCot and Kit Replacement Head Case Assembly It is designed to assist with the thermoregulation of infants weighing up to 10kg and is used by labor and delivery rooms and pediatric intensive care units. Infant Warmers Model Number (REF) IW910XXX, IW920XXX, IW931XXX, IW932XXX, IW933XXX, IW934XXX, IW951XXX, IW952XXX, IW953XXX, IW954XXX, IW980XXX, IW990XXX Spare Parts Model Number (REF) 043041129, 043041130, 043041131, 043042359, 648040142

Reason

Breakage of the nut that secures the heater head in place, which can cause the heater head to become partially detached and swing towards the bassinet. There is the potential that a heater head could contact a patient situated on the bassinet and cause a serious injury.

Identifiers

code_infoAffected Serial (SN) and Lot (LOT) Numbers Infant Warmers: 071116ZZZZZZ - 090810ZZZZZZ Spare Parts: 071116 - 090810