FDA Device recall Z-0210-2014
Composite Series Skull Clamp, one clamp per tote
- FDA Device
- Class II
- terminated
- Published 2013-11-20
On 2013-11-20 the FDA classified a Class II recall of Composite Series Skull Clamp, one clamp per tote by Integra Lifesciences, citing an investigation of an adverse trend of complaints for the Skull Clamp index locking knob not easily unlocking or disengaging and malfunctions of the 80 lbs. torque knob was initi….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-0210-2014 |
| Classification | Class II |
| Status | terminated |
| Initiated | 2013-08-29 |
| Published | 2013-11-20 |
| Last updated | 2026-09-13 11:53 UTC |
| Company | Integra Lifesciences |
| Distribution | AL, AR, CA, CT, FL, GA, IA, IL, MS, NY, OH, SC, TN, TX |
Product
Composite Series Skull Clamp, one clamp per tote. The MAYFIELD¿ Skull Clamp (A3059) is a cranial stabilization device, designed to provide rigid skeletal fixation. The Skull Clamp is placed on the patient's skull to hold their head and neck securely in a particular position when rigid fixation is desired. The clamp is indicated for use in open and percutaneous craniotomies as well as spinal surgery when rigid skeletal fixation is necessary
Reason
An investigation of an adverse trend of complaints for the Skull Clamp index locking knob not easily unlocking or disengaging and malfunctions of the 80 lbs. torque knob was initiated. None of the complaints resulted in a report of patient injury, adverse health consequence, or a delay / prolongation of the surgical procedure.
Identifiers
| code_info | The following serial numbers are affected by this recall: SS120076, SS120082, SS120083, SS120084, SS120088, SS120089, S…The following serial numbers are affected by this recall: SS120076, SS120082, SS120083, SS120084, SS120088, SS120089, SS120090, SS120092, SS120093, SS120094, SS120095, SS120111, SS120112, SS120113, SS120114, SS120115, SS120116, SS120117, SS120118, SS120119, SS120120, SS120121, SS120122, SS120123, SS120124, SS120125, SS120126, SS120127, SS120128, SS120130, SS120131, SS120133, SS120134, SS120135, SS120171, SS120252, SS120256, SS120257 |
|---|
Status history
No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:53 UTC. Current status: terminated.
Source
Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-0210-2014%22