Alloclassic SL Shaft

On 2012-11-14 the FDA classified a Class II recall of Alloclassic SL Shaft by Zimmer, citing zimmer Inc., is initiating a correction to the patient labels of products manufactured before March 2010 by the Zimmer GmbH production site in Winterthur, Switzerland, after recei….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-0210-2013
ClassificationClass II
Statusterminated
Initiated2012-09-04
Published2012-11-14
Last updated2026-09-13 11:53 UTC
CompanyZimmer
DistributionUS

Product

Alloclassic SL Shaft

Reason

Zimmer Inc., is initiating a correction to the patient labels of products manufactured before March 2010 by the Zimmer GmbH production site in Winterthur, Switzerland, after receiving four complaints of a discrepancy between product patient labels and product labels with respect to the reference and lot numbers of the product. While the information printed on the product label is correct, the info

Identifiers

code_info
Item 2839, 2840, 2841, 2842, 2843, 2844, 2845, 2846, 2847, 2848, 2949, 2851, 2852, 2883, 2884, 2885, 2886, and 2887 L…Item 2839, 2840, 2841, 2842, 2843, 2844, 2845, 2846, 2847, 2848, 2949, 2851, 2852, 2883, 2884, 2885, 2886, and 2887 Lots: 2359249, 2359251, 2359251, 2368739, 2413914, 2413914, 2426010, 2526138, 2335608, 2381324, 2387449, 2388541, 2401237, 2401237, 2492774, 2523553, 2355767, 2355767, 2361511, 2361511, 2371517, 2376476, 2440672, 2453100, 2479736, 2498363, 2529208, 2303302, 2347856, 2355765, 2355765, 2428247, 2428247, 2446699, 2456353, 2497037, 2506067, 2506067, 2509374, 2517494, 2316056, 2326792, 2335862, 2355428, 2370478, 2376478, 2382854, 2388539, 2391204, 2428742, 2431991, 2439733, 2453093, 2301633, 2343382, 2363829, 2363829, 2363829, 2370490, 2492887, 2513725, 2513725, 2513725, 2515404, 2515404, 2517796, 2529231, 2319316, 2340681, 2345324, 2359179, 2359179, 2413790, 2432646, 2479750, 2501280, 2509376, 2509376, 2509378, 2510392, 2510392, 2303201, 2333704, 2340906, 2368928, 2375002, 2388530, 2403651, 2445551, 2471087, 2507375, 2511623, 2523569, 2345382, 2396573, 2428657, 2432681, 246

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:53 UTC. Current status: terminated.

Source

Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-0210-2013%22

RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.