RecallRadar

FDA Device recall Z-0207-2025

Diasol Acid Concentrate REF G100325-10Dex 100

On 2024-10-30 the FDA classified a Class II recall of Diasol Acid Concentrate REF G100325-10Dex 100 by Diasol, citing due to labeling issue. The box label is different that the gallon label in that gallon label shows a lower Calcium concentration than the box label.

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-0207-2025
ClassificationClass II
Statusongoing
Initiated2022-07-27
Published2024-10-30
Last updated2026-09-13 11:57 UTC
CompanyDiasol
DistributionUS

Product

Diasol Acid Concentrate REF G100325-10Dex 100

Reason

Due to labeling issue. The box label is different that the gallon label in that gallon label shows a lower Calcium concentration than the box label.

Identifiers

code_infoCatalog Number: G100325-10Dex100 UDI code: B467100325101004 Lot Number:PHS06011/061722

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:57 UTC. Current status: ongoing.

Source

Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-0207-2025%22