RecallRadar

FDA Device recall Z-0207-2021

13 IN(33cm) APPX 1.0 ml, SMALLBORE SET, 4 MicroClave Clear, NanoClave T-Conn, Filter

On 2020-10-28 the FDA classified a Class II recall of 13 IN(33cm) APPX 1.0 ml, SMALLBORE SET, 4 MicroClave Clear, NanoClave T-Conn, Filter by Icu Medical, citing identification of a potential manufacturing defect on the internal surface of the NanoClave within specific lots of NanoClave sets, which may inhibit a proper seal with the NanoCl….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-0207-2021
ClassificationClass II
Statusterminated
Initiated2020-07-06
Published2020-10-28
Last updated2026-09-13 11:55 UTC
CompanyIcu Medical
DistributionUS

Product

13 IN(33cm) APPX 1.0 ml, SMALLBORE SET, 4 MicroClave Clear, NanoClave T-Conn, Filter; 1 unit per pouch, 50 pouches per case. UDI:(01)10887709085210(17)250101(30)50(10)4572801; (01)10887709085227(17)250101(30)50(10)4589920; (01)10887709085227(17)250101(30)50(10)4572950. The Intravascular Administration Set is a single use, sterile, non-pyrogenic device which provides access for the administration of fluids from a container to a patient's vascular system through the administration set's needle or catheter (which is inserted into a vein).

Reason

Identification of a potential manufacturing defect on the internal surface of the NanoClave within specific lots of NanoClave sets, which may inhibit a proper seal with the NanoClave spike.

Identifiers

code_infoLot Numbers:4572801, 4589920, 4572950.

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:55 UTC. Current status: terminated.