FDA Device recall Z-0205-2013
Vector TAS Modular Driver, a manual driver for Vector TAS dental screws
- FDA Device
- Class II
- terminated
- Published 2012-11-14
On 2012-11-14 the FDA classified a Class II recall of Vector TAS Modular Driver, a manual driver for Vector TAS dental screws by Ormco Sybronendo, citing the Instructions For Use (IFUs) for Vector TAS Modular Driver for some foreign language sections incorrectly lists the sterilization temperature as 130¿C. The correct sterilizatio….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-0205-2013 |
| Classification | Class II |
| Status | terminated |
| Initiated | 2011-12-12 |
| Published | 2012-11-14 |
| Last updated | 2026-09-13 11:53 UTC |
| Company | Ormco Sybronendo |
| Distribution | US |
Product
Vector TAS Modular Driver, a manual driver for Vector TAS dental screws.
Reason
The Instructions For Use (IFUs) for Vector TAS Modular Driver for some foreign language sections incorrectly lists the sterilization temperature as 130¿C. The correct sterilization temperature should be listed as 132¿C and/or 270¿F.
Identifiers
| code_info | Part Numbers 601-0007 and 601-0010, all lots |
|---|