FDA Device recall Z-0201-2014
Titan 3T MRI System, MRT-3010A/5. Product Usage: Usage: MRI system
- FDA Device
- Class II
- terminated
- Published 2013-11-20
On 2013-11-20 the FDA classified a Class II recall of Titan 3T MRI System, MRT-3010A/5. Product Usage: Usage: MRI system by Toshiba American Medical Systems, citing toshiba America Medical Systems, Inc. is recalling due to a potential problem with Specific Absorption Rate (SAR) values when using the Toshiba Titan 3T System. When the system is….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-0201-2014 |
| Classification | Class II |
| Status | terminated |
| Initiated | 2013-10-28 |
| Published | 2013-11-20 |
| Last updated | 2026-09-13 11:53 UTC |
| Company | Toshiba American Medical Systems |
| Distribution | US |
Product
Titan 3T MRI System, MRT-3010A/5. Product Usage: Usage: MRI system.
Reason
Toshiba America Medical Systems, Inc. is recalling due to a potential problem with Specific Absorption Rate (SAR) values when using the Toshiba Titan 3T System. When the system is used in combination with either the QD Head Coil (MJQH-142A) or the Large Knee SPEEDER (MJAJ-182A), SAR values lower than the actual SAR values may be displayed on the operating monitor.
Identifiers
| code_info | Lot numbers not utilized. Serial Numbers:A5B1192003 A5B1192004, A5B11Y2005, A5B11Z2007, A5C1212008, A5C1212009, A…Lot numbers not utilized. Serial Numbers:A5B1192003 A5B1192004, A5B11Y2005, A5B11Z2007, A5C1212008, A5C1212009, A5C1222010, A5C1232011, A5C1252012, A5C1262013, A5C1282014, A5C1292015, A5C12X2016, A5C12Y2017, A5C12Z2018, A5D1312019, A5D1322020, A5D1362021, XIB-A5A1082001, XIB-A5B11Y2006, A5D1372022, A5B1182002. |
|---|