FDA Device recall Z-0199-2025
medex TranStar KIDS KIT 34IN. CLOSED BLOOD SAMPLE KIT with 30cc/hr Flush Device 10/EA, Product Code REF MX9545
- FDA Device
- Class II
- ongoing
- Published 2024-10-30
On 2024-10-30 the FDA classified a Class II recall of medex TranStar KIDS KIT 34IN. CLOSED BLOOD SAMPLE KIT with 30cc/hr Flush Device 10/EA, Product Code REF MX9545 by Smiths Medical Asd, citing due to a manufacturing defect, users may experience inaccurate pressure monitoring, an interruption in pressure monitoring, or an inability to zero the device due to either a leak….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-0199-2025 |
| Classification | Class II |
| Status | ongoing |
| Initiated | 2024-09-24 |
| Published | 2024-10-30 |
| Last updated | 2026-09-13 11:57 UTC |
| Company | Smiths Medical Asd |
| Distribution | n/a |
Product
medex TranStar KIDS KIT 34IN. CLOSED BLOOD SAMPLE KIT with 30cc/hr Flush Device 10/EA, Product Code REF MX9545
Reason
Due to a manufacturing defect, users may experience inaccurate pressure monitoring, an interruption in pressure monitoring, or an inability to zero the device due to either a leak within the TranStar Disposable Pressure Transducer or from a defect within the TranStar Disposable Pressure Transducer chip.
Identifiers
| code_info | UDI/DI 50351688503068 (case), 10351688503060 (pouch), Lot Numbers: 4323136 |
|---|
Status history
No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:57 UTC. Current status: ongoing.
Other recalls by Smiths Medical Asd
- CADD Solis HSPCA Pump ***Updated 5/27/26*** Scope updated to apply specifically to customers using the wireless system
FDA Device · Class I · 2025-05-14 · Smiths Medical Asd - CADD-Solis VIP Ambulatory Infusion Pump
FDA Device · Class I · 2025-05-14 · Smiths Medical Asd - CADD-Solis Infusion Pump Models: 1) PUMP KIT, CADD-SOLIS, MDL 2110, V4.2, CE ENGLISH 1/EA, Model Number: 21-2111-0402-51
FDA Device · Class I · 2025-05-14 · Smiths Medical Asd - CADD Solis VIP Ambulatory Infusion Pump Models: 1) PUMP KIT, CADD-SOLIS VIP, MDL 2120, NA ENGLISH, MANUAL MODE, 1/EA, Model Number: 21-2127…
FDA Device · Class I · 2025-05-14 · Smiths Medical Asd - CADD-Solis HPCA Ambulatory Infusion Pump
FDA Device · Class I · 2025-05-14 · Smiths Medical Asd
All Smiths Medical Asd recalls
Source
Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-0199-2025%22
RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.