RecallRadar

FDA Device recall Z-0197-2015

Integra Titan Reverse Shoulder System right and left head cutting templates, 142.5 degrees

On 2014-11-19 the FDA classified a Class II recall of Integra Titan Reverse Shoulder System right and left head cutting templates, 142.5 degrees by Integra Lifesciences, citing A single lot of left and a single lot of right Reverse Shoulder System cutting templates were manufactured incorrectly. Specifically, the threaded handle has been welded backwards….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-0197-2015
ClassificationClass II
Statusterminated
Initiated2014-10-24
Published2014-11-19
Last updated2026-09-13 11:53 UTC
CompanyIntegra Lifesciences
DistributionUS

Product

Integra Titan Reverse Shoulder System right and left head cutting templates, 142.5 degrees. The Reverse Shoulder System Cutting Template, 142.5 degrees is an accessory for use with the Integra Titan Reverse Shoulder System. It is used to project the osteotomy angle on the humeral head.

Reason

A single lot of left and a single lot of right Reverse Shoulder System cutting templates were manufactured incorrectly. Specifically, the threaded handle has been welded backwards on the template resulting in the cutting angle being the reverse of what it should be.

Identifiers

code_infoLot Numbers PM0278 and PM0277