RecallRadar

FDA Device recall Z-0194-2024

EVIS EXERA III BRONCHOVIDEOSCOPE OLYMPUS Model Number BF-P190

On 2023-11-15 the FDA classified a Class I recall of EVIS EXERA III BRONCHOVIDEOSCOPE OLYMPUS Model Number BF-P190 by Olympus, citing there is a risk of endobronchial combustion if high-frequency cauterization is performed while supplying oxygen [and/or] the electrode section of the electrosurgical accessory is….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-0194-2024
ClassificationClass I
Statusongoing
Initiated2023-10-12
Published2023-11-15
Last updated2026-09-13 11:56 UTC
CompanyOlympus
DistributionUS

Product

EVIS EXERA III BRONCHOVIDEOSCOPE OLYMPUS Model Number BF-P190

Reason

There is a risk of endobronchial combustion if high-frequency cauterization is performed while supplying oxygen [and/or] the electrode section of the electrosurgical accessory is too close to the distal end of the endoscope.

Identifiers

code_infoModel BF-P190 UDI-DI: 4953170342110

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:56 UTC. Current status: ongoing.

Source

Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-0194-2024%22

RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.