FDA Device recall Z-0193-2024

EVIS EXERA b BRONCHOVIDEOSCOPE OLYMPUS Model Number BF-XT190

On 2023-11-15 the FDA classified a Class I recall of EVIS EXERA b BRONCHOVIDEOSCOPE OLYMPUS Model Number BF-XT190 by Olympus, citing there is a risk of endobronchial combustion if high-frequency cauterization is performed while supplying oxygen [and/or] the electrode section of the electrosurgical accessory is….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-0193-2024
ClassificationClass I
Statusongoing
Initiated2023-10-12
Published2023-11-15
Last updated2026-09-13 11:56 UTC
CompanyOlympus
DistributionUS

Product

EVIS EXERA b BRONCHOVIDEOSCOPE OLYMPUS Model Number BF-XT190

Reason

There is a risk of endobronchial combustion if high-frequency cauterization is performed while supplying oxygen [and/or] the electrode section of the electrosurgical accessory is too close to the distal end of the endoscope.

Identifiers

code_infoModel BF-XT190 UDI-DI: 4953170402470

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:56 UTC. Current status: ongoing.