RecallRadar

FDA Device recall Z-0193-2023

Insulin Pump/Model: MiniMed 630G/ MMT-1715, MMT-1755, MMT-1754

On 2022-11-16 the FDA classified a Class II recall of Insulin Pump/Model: MiniMed 630G/ MMT-1715, MMT-1755, MMT-1754 by Medtronic Minimed, citing medtronic identified a cybersecurity vulnerability in the MiniMed 600 series Insulin Pump Systems associated with the communication protocol that could allow unauthorized access t….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-0193-2023
ClassificationClass II
Statusongoing
Initiated2022-09-20
Published2022-11-16
Last updated2026-09-13 11:56 UTC
CompanyMedtronic Minimed
DistributionAK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, PR, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, WY

Product

Insulin Pump/Model: MiniMed 630G/ MMT-1715, MMT-1755, MMT-1754; MiniMed 670G/ MMT-1780, MMT-1781, MMT-1782, MMT-1760, MMT-1761, MMT-1762, MMT-1740, MMT-1741, MMT-1742

Reason

Medtronic identified a cybersecurity vulnerability in the MiniMed 600 series Insulin Pump Systems associated with the communication protocol that could allow unauthorized access to the pump system. This unauthorized access could be used to deliver too much or too little insulin through delivery of an unintended insulin bolus or because insulin delivery is slowed or stopped, which could lead to hypoglycemia or hyperglycemia. The Remote Bolus feature on insulin pumps should be turned off to prevent the unauthorized access.

Identifiers

code_info
Model/UDI-GTIN(All Serial Numbers): MMT-1715/643000000000, 763000072520, 643169873834, 00643169656840; MMT-1755/6431697…Model/UDI-GTIN(All Serial Numbers): MMT-1715/643000000000, 763000072520, 643169873834, 00643169656840; MMT-1755/643169752726, 643169873841, 643169878648, 763000179632, 00643169752726; MMT-1754/643169976344; MMT-1780/643169949706, 643169939202, 763000072537, 643000000000, 00643169939219; MMT-1781/763000187415, 763000283568, 00763000140601, 00763000225575, 00763000235765, 00763000187415, 00763000283551, 00763000240707; MMT-1782/763000187422, 763000283599, 00763000140618, 00763000225612, 00763000235772, 00763000187422, 00643169957640, 00763000283582; MMT-1760/643169946040, 763000278427, 763000179649, 643169946057, 763000367084, 00643169946057; MMT-1761/00763000367657, 00763000165413; MMT-1762/00763000239404, 00763000203313, 00763000216481, 00763000187408, 00763000165475; MMT-1740/Investigational; MMT-1741/00763000303556, 00763000408916; MMT-1742/00763000250232

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:56 UTC. Current status: ongoing.

Source

Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-0193-2023%22

RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.