RecallRadar

FDA Device recall Z-0192-2024

EVIS EXERA III BRONCHOVIDEOSCOPE Model Numbers BF-1TH190, BF-H190, BF-Q190 and BRONCHOVIDEOSCOPE OLYMPUS BF-Q170 Model Number BF-Q170

On 2023-11-15 the FDA classified a Class I recall of EVIS EXERA III BRONCHOVIDEOSCOPE Model Numbers BF-1TH190, BF-H190, BF-Q190 and BRONCHOVIDEOSCOPE OLYMPUS BF-Q170 Model Number BF-Q170 by Olympus, citing there is a risk of endobronchial combustion if high-frequency cauterization is performed while supplying oxygen [and/or] the electrode section of the electrosurgical accessory is….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-0192-2024
ClassificationClass I
Statusongoing
Initiated2023-10-12
Published2023-11-15
Last updated2026-09-13 11:56 UTC
CompanyOlympus
DistributionUS

Product

EVIS EXERA III BRONCHOVIDEOSCOPE Model Numbers BF-1TH190, BF-H190, BF-Q190 and BRONCHOVIDEOSCOPE OLYMPUS BF-Q170 Model Number BF-Q170

Reason

There is a risk of endobronchial combustion if high-frequency cauterization is performed while supplying oxygen [and/or] the electrode section of the electrosurgical accessory is too close to the distal end of the endoscope.

Identifiers

code_infoModel BF-1TH190 UDI-DI: 4953170335181; Model BF-H190 UDI-DI: 4953170335174; Model BF-Q170 UDI-DI: 4953170342912 Model BF-Q190 UDI-DI: 4953170335198

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:56 UTC. Current status: ongoing.

Source

Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-0192-2024%22